Home > Analytical Instruments > Water Activity Meter > Water Activity Meter
Water activity meter AW01-MC is based on the mirror condensation dew point method.
Widely used in pharmaceuticals, food, and other fields to ensure that the risk of microbial contamination in products is controllable.
The dew point temperature is determined by cooling a mirror in the sample chamber until condensation occurs and recording the mirror temperature at that point. The saturated vapor pressure (P) can be obtained from a table using this dew point temperature. The ratio of this saturated vapor pressure (P) to the saturated vapor pressure (P0) of pure water is the water activity (aw=P/P0).
| Test Range: | (0.030-1.000)aw |
| Max. Allowable Error: | ±0.003aw |
| Resolution: | ±0.0001aw |
| Repeatability Error: | ≤3% |
| Test Time: | 6-8min |
| Temperature Control Range: | (10-50)℃ |
| Temperature Control Accuracy: | ±0.2℃ |
| Environment: | Temperature (4-50)℃, (10%-85%)RH (non-condensing) |
| Sample Temperature: | (10-50)℃ accuracy ±0.2℃, resolution 0.01℃ |
| Input Power: | (100~240)±10%VAC, 50/60Hz |
| Power: | ≤100W |
| Sample Cup Capacity: | 7mL |
| Data Storage: | More than ten years of storage period (depending on the size of the memory, each group includes water activity, temperature, time, date, operator and sensor type) |
It can be used in the pharmaceutical, food, and biodegradable industries to test drugs, food, soil, compost, etc.