Home > Pharma Processing > Sterile Transfer > Sterile Pass Box > Sterile Pass Box
The STP Series Sterile Pass Box achieve biological decontamination of the interior space,surfaces, and items (outer surfaces) through an integrated vaporized hydrogen peroxide (VH2O2) system. The integrated VH2O2 system is linked with the sterile pass box using SIEMENS PLC control. Communication between the sterile pass box and the VH2O2 system allows for the setting of relevant decontamination parameters, after which the system’s built-in program supports unattended automatic operation.
| Airflow: Turbulence | Cleanliness: GMP Grade A |
| Biological Decontamination Cycle Time: 150 minutes | Biological Decontamination Capability: Log reduction value≥6 |
| VH2O2 Residual Level: ≤1ppm | High-Efficiency Filter Grade: Supply air HEPA (H14), Exhaust air HEPA (H14) |
| Printing: Includes biological decontamination cycle parameter saving and printing functions | Alarm: High pressure/low pressure alarm, door open timeout alarm, door lock alarm, concentration/temperature alarm |
| Exhaust Interface: 3-inch or 4-inch sanitary interface | Hydrogen Peroxide Gas Decontamination Cycle Interface: 1.5-inch sanitary interface |
| Material of Inside Chamber: SUS316L stainless steel |
It is widely used in microelectronics, biological laboratories, pharmaceutical factories, hospitals, and the food industry—any environment requiring air purification. It is suitable for transferring various clean and dry items in sterile production, including the outer packaging of packaging materials, instruments, raw materials, accessories, and environmental monitoring equipment that enter A and B critical areas.